Classify medical devices under EU MDR 2017/745, build technical documentation, plan clinical evaluations, and manage post-market surveillance and EUDAMED integration.

...Expand all
22
Updated time August 29, 2026

MDR 2017/745 Specialist

EU MDR compliance patterns for medical device classification, technical documentation, and clinical evidence.

Table of Contents

  • Device Classification Workflow
  • Technical Documentation
  • Clinical Evidence
  • Post-Market Surveillance
  • EUDAMED and UDI
  • Reference Documentation
  • Tools

Device Classification Workflow

Classify device under MDR Annex VIII:

  1. Identify device duration (transient, short-term, long-term)
  2. Determine invasiveness level (non-invasive, body orifice, surgical)
  3. Assess body system contact (CNS, cardiac, other)
  4. Check if active device (energy dependent)
  5. Apply classification rules 1-22
  6. For software, apply MDCG 2019-11 algorithm
  7. Document classification rationale
  8. Validation: Classification confirmed with Notified Body

Classification Matrix

Factor Class I Class IIa Class IIb Class III
Duration Any Short-term Long-term Long-term
Invasiveness Non-invasive Body orifice Surgical Implantable
System Any Non-critical Critical organs CNS/cardiac
Risk Lowest Low-medium Medium-high Highest

Software Classification (MDCG 2019-11)

Information Use Condition Severity Class
Informs decision Non-serious IIa
Informs decision Serious IIb
Drives/treats Critical III

Classification Examples

Example 1: Absorbable Surgical Suture

  • Rule 8 (implantable, long-term)
  • Duration: > 30 days (absorbed)
  • Contact: General tissue
  • Classification: Class IIb

Example 2: AI Diagnostic Software

  • Rule 11 + MDCG 2019-11
  • Function: Diagnoses serious condition
  • Classification: Class IIb

Example 3: Cardiac Pacemaker

  • Rule 8 (implantable)
  • Contact: Central circulatory system
  • Classification: Class III

Technical Documentation

Prepare technical file per Annex II and III:

  1. Create device description (variants, accessories, intended purpose)
  2. Develop labeling (Article 13 requirements, IFU)
  3. Document design and manufacturing process
  4. Complete GSPR compliance matrix
  5. Prepare benefit-risk analysis
  6. Compile verification and validation evidence
  7. Integrate risk management file (ISO 14971)
  8. Validation: Technical file reviewed for completeness

Technical File Structure

ANNEX II TECHNICAL DOCUMENTATION
├── Device description and UDI-DI
├── Label and instructions for use
├── Design and manufacturing info
├── GSPR compliance matrix
├── Benefit-risk analysis
├── Verification and validation
└── Clinical evaluation report

GSPR Compliance Checklist

Requirement Evidence Status
Safe design (GSPR 1-3) Risk management file
Chemical properties (GSPR 10.1) Biocompatibility report
Infection risk (GSPR 10.2) Sterilization validation
Software requirements (GSPR 17) IEC 62304 documentation
Labeling (GSPR 23) Label artwork, IFU

Conformity Assessment Routes

Class Route NB Involvement
I Annex II self-declaration None
Is/Im Annex II + IX/XI Sterile/measuring aspects
IIa Annex II + IX or XI Product or QMS
IIb Annex IX, or Annex X + XI QMS + tech doc assessment, or type exam + production
III Annex IX, or Annex X + XI Full QMS + product dossier, or type exam + production

Clinical Evidence

Develop clinical evidence strategy per Annex XIV:

  1. Define clinical claims and endpoints
  2. Conduct systematic literature search
  3. Appraise clinical data quality
  4. Assess equivalence (technical, biological, clinical)
  5. Identify evidence gaps
  6. Determine if clinical investigation required
  7. Prepare Clinical Evaluation Report (CER)
  8. Validation: CER reviewed by qualified evaluator

Evidence Requirements by Class

Class Minimum Evidence Investigation
I Risk-benefit analysis Not typically required
IIa Literature + post-market May be required
IIb Systematic literature review Often required
III Comprehensive clinical data Required (Article 61)

Clinical Evaluation Report Structure

CER CONTENTS
├── Executive summary
├── Device scope and intended purpose
├── Clinical background (state of the art)
├── Literature search methodology
├── Data appraisal and analysis
├── Safety and performance conclusions
├── Benefit-risk determination
└── PMCF plan summary

Qualified Evaluator Requirements

  • Medical degree or equivalent healthcare qualification
  • 4+ years clinical experience in relevant field
  • Training in clinical evaluation methodology
  • Understanding of MDR requirements

Post-Market Surveillance

Establish PMS system per Chapter VII:

  1. Develop PMS plan (Article 84)
  2. Define data collection methods
  3. Establish complaint handling procedures
  4. Create vigilance reporting process
  5. Plan Periodic Safety Update Reports (PSUR)
  6. Integrate with PMCF activities
  7. Define trend analysis and signal detection
  8. Validation: PMS system audited annually

PMS System Components

Component Requirement Frequency
PMS Plan Article 84 Maintain current
PSUR Article 86 — Class IIa and higher Per Art. 86(1) schedule below
PMCF Plan Annex XIV Part B Update with CER
PMCF Report Annex XIV Part B Annual (Class III)
Vigilance Articles 87-92 As events occur

PSUR Schedule

Class Frequency (MDR Art. 86(1))
Class III Updated at least annually
Class IIb (all, incl. implantable) Updated at least annually
Class IIa When necessary, at least every 2 years
Class I No PSUR — PMS report instead (Art. 85)

Serious Incident Reporting

Timeline Requirement
2 days Serious public health threat
10 days Death or serious deterioration
15 days Other serious incidents

EUDAMED and UDI

Implement UDI system per Article 27:

  1. Obtain issuing entity code (GS1, HIBCC, ICCBBA)
  2. Assign UDI-DI to each device variant
  3. Assign UDI-PI (production identifier)
  4. Apply UDI carrier to labels (AIDC + HRI)
  5. Register actor in EUDAMED
  6. Register devices in EUDAMED
  7. Upload certificates when available
  8. Validation: UDI verified on sample labels

EUDAMED Modules

Module Content Actor
Actor Company registration Manufacturer, AR
UDI/Device Device and variant data Manufacturer
Certificates NB certificates Notified Body
Clinical Investigation Study registration Sponsor
Vigilance Incident reports Manufacturer
Market Surveillance Authority actions Competent Authority

UDI Label Requirements

Required elements per Article 13:

  • UDI-DI (device identifier)
  • UDI-PI (production identifier) for Class II+
  • AIDC format (barcode/RFID)
  • HRI format (human-readable)
  • Manufacturer name and address
  • Lot/serial number
  • Expiration date (if applicable)

Reference Documentation

MDR Classification Guide

references/mdr-classification-guide.md contains:

  • Complete Annex VIII classification rules (Rules 1-22)
  • Software classification per MDCG 2019-11
  • Worked classification examples
  • Conformity assessment route selection

Clinical Evidence Requirements

references/clinical-evidence-requirements.md contains:

  • Clinical evidence framework and hierarchy
  • Literature search methodology
  • Clinical Evaluation Report structure
  • PMCF plan and evaluation report guidance

Technical Documentation Templates

references/technical-documentation-templates.md contains:

  • Annex II and III content requirements
  • Design History File structure
  • GSPR compliance matrix template
  • Declaration of Conformity template
  • Notified Body submission checklist

Tools

MDR Gap Analyzer

# Quick gap analysis
python scripts/mdr_gap_analyzer.py --device "Device Name" --class IIa

# JSON output for integration
python scripts/mdr_gap_analyzer.py --device "Device Name" --class III --output json

# Interactive assessment
python scripts/mdr_gap_analyzer.py --interactive

Analyzes device against MDR requirements, identifies compliance gaps, generates prioritized recommendations.

Output includes:

  • Requirements checklist by category
  • Gap identification with priorities
  • Critical gap highlighting
  • Compliance roadmap recommendations

Notified Body Interface

Selection Criteria

Factor Considerations
Designation scope Covers your device type
Capacity Timeline for initial audit
Geographic reach Markets you need to access
Technical expertise Experience with your technology
Fee structure Transparency, predictability

Pre-Submission Checklist

  • Technical documentation complete
  • GSPR matrix fully addressed
  • Risk management file current
  • Clinical evaluation report complete
  • QMS (ISO 13485) certified
  • Labeling and IFU finalized
  • Validation: Internal gap assessment complete
View on GitHub
---
name: mdr-745-specialist
description: Classify medical devices under EU MDR 2017/745, build technical documentation, plan clinical evaluations, and manage post-market surveillance and EUDAMED integration.
---

# MDR 2017/745 Specialist

EU MDR compliance patterns for medical device classification, technical documentation, and clinical evidence.

---

## Table of Contents

- [Device Classification Workflow](#device-classification-workflow)
- [Technical Documentation](#technical-documentation)
- [Clinical Evidence](#clinical-evidence)
- [Post-Market Surveillance](#post-market-surveillance)
- [EUDAMED and UDI](#eudamed-and-udi)
- [Reference Documentation](#reference-documentation)
- [Tools](#tools)

---

## Device Classification Workflow

Classify device under MDR Annex VIII:

1. Identify device duration (transient, short-term, long-term)
2. Determine invasiveness level (non-invasive, body orifice, surgical)
3. Assess body system contact (CNS, cardiac, other)
4. Check if active device (energy dependent)
5. Apply classification rules 1-22
6. For software, apply MDCG 2019-11 algorithm
7. Document classification rationale
8. **Validation:** Classification confirmed with Notified Body

### Classification Matrix

| Factor | Class I | Class IIa | Class IIb | Class III |
|--------|---------|-----------|-----------|-----------|
| Duration | Any | Short-term | Long-term | Long-term |
| Invasiveness | Non-invasive | Body orifice | Surgical | Implantable |
| System | Any | Non-critical | Critical organs | CNS/cardiac |
| Risk | Lowest | Low-medium | Medium-high | Highest |

### Software Classification (MDCG 2019-11)

| Information Use | Condition Severity | Class |
|-----------------|-------------------|-------|
| Informs decision | Non-serious | IIa |
| Informs decision | Serious | IIb |
| Drives/treats | Critical | III |

### Classification Examples

**Example 1: Absorbable Surgical Suture**
- Rule 8 (implantable, long-term)
- Duration: > 30 days (absorbed)
- Contact: General tissue
- Classification: **Class IIb**

**Example 2: AI Diagnostic Software**
- Rule 11 + MDCG 2019-11
- Function: Diagnoses serious condition
- Classification: **Class IIb**

**Example 3: Cardiac Pacemaker**
- Rule 8 (implantable)
- Contact: Central circulatory system
- Classification: **Class III**

---

## Technical Documentation

Prepare technical file per Annex II and III:

1. Create device description (variants, accessories, intended purpose)
2. Develop labeling (Article 13 requirements, IFU)
3. Document design and manufacturing process
4. Complete GSPR compliance matrix
5. Prepare benefit-risk analysis
6. Compile verification and validation evidence
7. Integrate risk management file (ISO 14971)
8. **Validation:** Technical file reviewed for completeness

### Technical File Structure

```
ANNEX II TECHNICAL DOCUMENTATION
├── Device description and UDI-DI
├── Label and instructions for use
├── Design and manufacturing info
├── GSPR compliance matrix
├── Benefit-risk analysis
├── Verification and validation
└── Clinical evaluation report
```

### GSPR Compliance Checklist

| Requirement | Evidence | Status |
|-------------|----------|--------|
| Safe design (GSPR 1-3) | Risk management file | ☐ |
| Chemical properties (GSPR 10.1) | Biocompatibility report | ☐ |
| Infection risk (GSPR 10.2) | Sterilization validation | ☐ |
| Software requirements (GSPR 17) | IEC 62304 documentation | ☐ |
| Labeling (GSPR 23) | Label artwork, IFU | ☐ |

### Conformity Assessment Routes

| Class | Route | NB Involvement |
|-------|-------|----------------|
| I | Annex II self-declaration | None |
| Is/Im | Annex II + IX/XI | Sterile/measuring aspects |
| IIa | Annex II + IX or XI | Product or QMS |
| IIb | Annex IX, or Annex X + XI | QMS + tech doc assessment, or type exam + production |
| III | Annex IX, or Annex X + XI | Full QMS + product dossier, or type exam + production |

---

## Clinical Evidence

Develop clinical evidence strategy per Annex XIV:

1. Define clinical claims and endpoints
2. Conduct systematic literature search
3. Appraise clinical data quality
4. Assess equivalence (technical, biological, clinical)
5. Identify evidence gaps
6. Determine if clinical investigation required
7. Prepare Clinical Evaluation Report (CER)
8. **Validation:** CER reviewed by qualified evaluator

### Evidence Requirements by Class

| Class | Minimum Evidence | Investigation |
|-------|------------------|---------------|
| I | Risk-benefit analysis | Not typically required |
| IIa | Literature + post-market | May be required |
| IIb | Systematic literature review | Often required |
| III | Comprehensive clinical data | Required (Article 61) |

### Clinical Evaluation Report Structure

```
CER CONTENTS
├── Executive summary
├── Device scope and intended purpose
├── Clinical background (state of the art)
├── Literature search methodology
├── Data appraisal and analysis
├── Safety and performance conclusions
├── Benefit-risk determination
└── PMCF plan summary
```

### Qualified Evaluator Requirements

- Medical degree or equivalent healthcare qualification
- 4+ years clinical experience in relevant field
- Training in clinical evaluation methodology
- Understanding of MDR requirements

---

## Post-Market Surveillance

Establish PMS system per Chapter VII:

1. Develop PMS plan (Article 84)
2. Define data collection methods
3. Establish complaint handling procedures
4. Create vigilance reporting process
5. Plan Periodic Safety Update Reports (PSUR)
6. Integrate with PMCF activities
7. Define trend analysis and signal detection
8. **Validation:** PMS system audited annually

### PMS System Components

| Component | Requirement | Frequency |
|-----------|-------------|-----------|
| PMS Plan | Article 84 | Maintain current |
| PSUR | Article 86 — Class IIa and higher | Per Art. 86(1) schedule below |
| PMCF Plan | Annex XIV Part B | Update with CER |
| PMCF Report | Annex XIV Part B | Annual (Class III) |
| Vigilance | Articles 87-92 | As events occur |

### PSUR Schedule

| Class | Frequency (MDR Art. 86(1)) |
|-------|-----------------------------|
| Class III | Updated at least annually |
| Class IIb (all, incl. implantable) | Updated at least annually |
| Class IIa | When necessary, at least every 2 years |
| Class I | No PSUR — PMS report instead (Art. 85) |

### Serious Incident Reporting

| Timeline | Requirement |
|----------|-------------|
| 2 days | Serious public health threat |
| 10 days | Death or serious deterioration |
| 15 days | Other serious incidents |

---

## EUDAMED and UDI

Implement UDI system per Article 27:

1. Obtain issuing entity code (GS1, HIBCC, ICCBBA)
2. Assign UDI-DI to each device variant
3. Assign UDI-PI (production identifier)
4. Apply UDI carrier to labels (AIDC + HRI)
5. Register actor in EUDAMED
6. Register devices in EUDAMED
7. Upload certificates when available
8. **Validation:** UDI verified on sample labels

### EUDAMED Modules

| Module | Content | Actor |
|--------|---------|-------|
| Actor | Company registration | Manufacturer, AR |
| UDI/Device | Device and variant data | Manufacturer |
| Certificates | NB certificates | Notified Body |
| Clinical Investigation | Study registration | Sponsor |
| Vigilance | Incident reports | Manufacturer |
| Market Surveillance | Authority actions | Competent Authority |

### UDI Label Requirements

Required elements per Article 13:

- [ ] UDI-DI (device identifier)
- [ ] UDI-PI (production identifier) for Class II+
- [ ] AIDC format (barcode/RFID)
- [ ] HRI format (human-readable)
- [ ] Manufacturer name and address
- [ ] Lot/serial number
- [ ] Expiration date (if applicable)

---

## Reference Documentation

### MDR Classification Guide

`references/mdr-classification-guide.md` contains:

- Complete Annex VIII classification rules (Rules 1-22)
- Software classification per MDCG 2019-11
- Worked classification examples
- Conformity assessment route selection

### Clinical Evidence Requirements

`references/clinical-evidence-requirements.md` contains:

- Clinical evidence framework and hierarchy
- Literature search methodology
- Clinical Evaluation Report structure
- PMCF plan and evaluation report guidance

### Technical Documentation Templates

`references/technical-documentation-templates.md` contains:

- Annex II and III content requirements
- Design History File structure
- GSPR compliance matrix template
- Declaration of Conformity template
- Notified Body submission checklist

---

## Tools

### MDR Gap Analyzer

```bash
# Quick gap analysis
python scripts/mdr_gap_analyzer.py --device "Device Name" --class IIa

# JSON output for integration
python scripts/mdr_gap_analyzer.py --device "Device Name" --class III --output json

# Interactive assessment
python scripts/mdr_gap_analyzer.py --interactive
```

Analyzes device against MDR requirements, identifies compliance gaps, generates prioritized recommendations.

**Output includes:**
- Requirements checklist by category
- Gap identification with priorities
- Critical gap highlighting
- Compliance roadmap recommendations

---

## Notified Body Interface

### Selection Criteria

| Factor | Considerations |
|--------|----------------|
| Designation scope | Covers your device type |
| Capacity | Timeline for initial audit |
| Geographic reach | Markets you need to access |
| Technical expertise | Experience with your technology |
| Fee structure | Transparency, predictability |

### Pre-Submission Checklist

- [ ] Technical documentation complete
- [ ] GSPR matrix fully addressed
- [ ] Risk management file current
- [ ] Clinical evaluation report complete
- [ ] QMS (ISO 13485) certified
- [ ] Labeling and IFU finalized
- [ ] **Validation:** Internal gap assessment complete

All Files

0 files

Install mdr-745-specialist

Download and extract the skill files to your .claude/skills/ directory.

Download ZIP

Clone the repository and copy the skill files to your project.

git clone https://github.com/alirezarezvani/claude-skills/tree/main/ra-qm-team/skills/mdr-745-specialist # Copy SKILL.md to your .claude/skills/ directory

Copy Copy
Quick Setup: Copy the skill folder to .claude/skills/ Claude will automatically detect and use the skill

Related Skills

golang-dependency-injection
Updated time June 29, 2026
nuxthub
Updated time August 23, 2026
tc-tracker
Updated time August 27, 2026
code-quality
Updated time August 22, 2026
OR